E498/E498M-11 Standard Practice for Leaks Using the Mass Spectrometer Leak Detector or Residual Gas Analyzer in the Tracer Probe Mode
Published By | Publication Date | Number of Pages |
ASTM | 2011 | 4 |
ASTM E498/E498M-11
Historical Standard: Standard Practice for Leaks Using the Mass Spectrometer Leak Detector or Residual Gas Analyzer in the Tracer Probe Mode
ASTM E498/E498M
Scope
1.1 This practice covers procedures for testing and locating the sources of gas leaking at the rate of 1 × 10−8 Pa m3/s (1 × 10−9 Std cm 3/s) or greater. The test may be conducted on any object to be tested that can be evacuated and to the other side of which helium or other tracer gas may be applied.
1.2 Three test methods are described:
1.2.1 Test Method A—For the object under test capable of being evacuated, but having no inherent pumping capability.
1.2.2 Test Method B—For the object under test with integral pumping capability.
1.2.3 Test Method C—For the object under test as in Test Method B, in which the vacuum pumps of the object under test replace those normally used in the leak detector.
1.3 Units—The values stated in either SI or std-cc/sec units are to be regarded separately as standard. The values stated in each system may not be exact equivalents: therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Keywords
bell jar leak test; bomb mass spectrometer leak test; helium lead testing; helium leak test; leak testing; mass spectrometer leak testing; sealed object mass spectrometer leak test; Bell jar leak test; Bomb mass spectrometer leak test; Helium leak testing; Leak testing–mass spectrometer leak detector (MSLD); Mass spectrometer leak detector (MSLD); Probe methods; Residual gas analyzer (RGA); Sealed object mass spectrometer leak test; Tracer probe mode; Vacuum testing
ICS Code
ICS Number Code 19.100 (Non-destructive testing)
DOI: 10.1520/E0498_E0498M-11
PDF Catalog
PDF Pages | PDF Title |
---|---|
1 | Scope Referenced Documents Terminology Summary of Practice |
2 | Personnel Qualification Significance and Use Interferences Apparatus Calibration of MSLD |
3 | Procedure Apparatus Calibration of MSLD Preparation of Apparatus Test Procedure Apparatus Preparation of Apparatus |
4 | Calibration Test Procedure Keywords |