Email:[email protected]

1.1 It is common practice for orthopaedic implant manufacturers to apply permanent identification to implant components. In this regard, Practice F 86 describes recommended locations and methods of marking for metallic implants.

1.2 The purpose of this practice is to ( 1 ) recommend that orthopaedic implants be permanently marked, and ( 2 ) recommend practical amounts of information that should be included in the marking. It is recognized, however, that marking is not practical in some cases (see 4.1).

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.