BS EN ISO 4037-2:2021
$167.15
Radiological protection. X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy – Dosimetry for radiation protection over the energy ranges from 8 keV to 1,3 MeV and 4 MeV to 9 MeV
Published By | Publication Date | Number of Pages |
BSI | 2021 | 36 |
This document specifies the procedures for the dosimetry of X and gamma reference radiation for the calibration of radiation protection instruments over the energy range from approximately 8 keV to 1,3 MeV and from 4 MeV to 9 MeV and for air kerma rates above 1 ĀµGy/h. The considered measuring quantities are the air kerma free-in-air, Ka, and the phantom related operational quantities of the International Commission on Radiation Units and Measurements (ICRU)[2], H*(10), Hp(10), H’(3), Hp(3), H’(0,07) and Hp(0,07), together with the respective dose rates. The methods of production are given in ISO 4037ā1.
This document can also be used for the radiation qualities specified in ISO 4037ā1:2019, Annexes A, B and C, but this does not mean that a calibration certificate for radiation qualities described in these annexes is in conformity with the requirements of ISO 4037.
The requirements and methods given in this document are targeted at an overall uncertainty (k = 2) of the dose(rate) of about 6 % to 10 % for the phantom related operational quantities in the reference fields. To achieve this, two production methods of the reference fields are proposed in ISO 4037ā1.
The first is to produce āmatched reference fieldsā, which follow the requirements so closely that recommended conversion coefficients can be used. The existence of only a small difference in the spectral distribution of the āmatched reference fieldā compared to the nominal reference field is validated by procedures, which are given and described in detail in this document. For matched reference radiation fields, recommended conversion coefficients are given in ISO 4037ā3 only for specified distances between source and dosemeter, e.g., 1,0 m and 2,5 m. For other distances, the user has to decide if these conversion coefficients can be used.
The second method is to produce ācharacterized reference fieldsā. Either this is done by determining the conversion coefficients using spectrometry, or the required value is measured directly using secondary standard dosimeters. This method applies to any radiation quality, for any measuring quantity and, if applicable, for any phantom and angle of radiation incidence. The conversion coefficients can be determined for any distance, provided the air kerma rate is not below 1 ĀµGy/h.
Both methods require charged particle equilibrium for the reference field. However this is not always established in the workplace field for which the dosemeter shall be calibrated. This is especially true at photon energies without inherent charged particle equilibrium at the reference depth d, which depends on the actual combination of energy and reference depth d. Electrons of energies above 65 keV, 0,75 MeV and 2,1 MeV can just penetrate 0,07 mm, 3 mm and 10 mm of ICRU tissue, respectively, and the radiation qualities with photon energies above these values are considered as radiation qualities without inherent charged particle equilibrium for the quantities defined at these depths.
This document is not applicable for the dosimetry of pulsed reference fields.
PDF Catalog
PDF Pages | PDF Title |
---|---|
2 | undefined |
7 | Foreword |
8 | Introduction |
9 | 1 Scope |
10 | 2 Normative references 3 Terms and definitions |
11 | 4 Standard instrument 4.1 General 4.2 Calibration of the standard instrument 4.3 Energy dependence of the response of the standard instrument |
12 | 5 Conversion from the measured quantity air kerma, Ka, to the required phantom related measuring quantity 5.1 General |
14 | 5.2 Determination of conversion coefficients 5.2.1 General 5.2.2 Calculation of conversion coefficients from spectral fluence |
15 | 5.3 Validation of reference fields and of listed conversion coefficients using dosimetry |
16 | 6 Direct calibration of the reference field in terms of the required phantom related measuring quantity 7 Measurement procedures applicable to ionization chambers 7.1 Geometrical conditions 7.2 Chamber support and stem scatter 7.3 Location and orientation of the standard chamber 7.4 Measurement corrections 7.4.1 General |
17 | 7.4.2 Corrections for air temperature, pressure and humidity variation from reference calibration conditions 7.4.3 Corrections for radiation-induced leakage, including ambient radiation |
18 | 7.4.4 Incomplete ion collection 7.4.5 Beam non-uniformity 8 Additional procedures and precautions specific to gamma radiation dosimetry using radionuclide sources 8.1 Use of certified source output 8.2 Use of electron equilibrium caps 8.3 Radioactive source decay 8.4 Radionuclide impurities 8.5 Interpolation between calibration positions |
19 | 9 Additional procedures and precautions specific to X-radiation dosimetry 9.1 Variation of X-radiation output 9.2 Monitor 9.3 Adjustment of air kerma rate |
20 | 10 Dosimetry of reference radiation at photon energies between 4 MeV and 9 MeV 10.1 Dosimetric quantities 10.2 Measurement of the dosimetric quantities 10.2.1 General |
21 | 10.2.2 Air kerma (rate) 10.2.3 Phantom related operational quantities H*(10), Hp(10), H'(3) and Hp(3) 10.3 Measurement geometry 10.4 Monitor |
22 | 10.5 Determination of air kerma (rate) free-in-air 10.5.1 General 10.5.2 Measurement conditions 10.5.3 Direct measurement with an ionization chamber |
25 | 10.5.4 Determination of air kerma (rate) from photon fluence (rate) |
26 | 11 Uncertainty of measurement 11.1 General 11.2 Components of uncertainty 11.2.1 General 11.2.2 Uncertainties in the calibration of a secondary standard |
27 | 11.2.3 Uncertainties in the measurements of the reference radiation due to the standard instrument and its use 11.3 Statement of uncertainty |
28 | Annex A (normative) Technical details of the instruments and their operation |
31 | Annex B (informative) Measurement of photon spectra |
34 | Bibliography |