BS ISO 16531:2013
$142.49
Surface chemical analysis. Depth profiling. Methods for ion beam alignment and the associated measurement of current or current density for depth profiling in AES and XPS
Published By | Publication Date | Number of Pages |
BSI | 2013 | 30 |
PDF Catalog
PDF Pages | PDF Title |
---|---|
6 | Foreword |
7 | Introduction |
9 | Section sec_1 Section sec_2 Section sec_3 1 Scope 2 Normative references 3 Terms, definitions, symbols and abbreviated terms |
10 | Section sec_4 Section sec_4.1 Section sec_4.2 4 System requirements 4.1 General 4.2 Limitations |
11 | Section sec_5 Section sec_5.1 Section sec_5.2 Section sec_5.2.1 5 Ion beam alignment methods 5.1 General 5.2 Important issues to be considered prior to ion beam alignment |
12 | Section sec_5.2.2 Figure fig_1 |
13 | Table tab_1 Section sec_5.2.3 |
14 | Section sec_5.2.4 Figure fig_2 Section sec_5.2.5 Section sec_5.3 Section sec_5.3.1 5.3 Alignment using circular-aperture Faraday cup |
15 | Section sec_5.3.2 Section sec_5.3.3 Section sec_5.3.4 Figure fig_3 Section sec_5.3.5 |
16 | Section sec_5.3.6 Figure fig_4 Section sec_5.3.7 |
17 | Section sec_5.3.8 Section sec_5.4 Section sec_5.5 Section sec_5.5.1 Section sec_5.5.2 Section sec_5.5.3 5.4 Alignment using elliptical-aperture Faraday cup 5.5 Alignment using images from ion-induced secondary electrons during ion beam rastering |
18 | Section sec_5.5.4 Section sec_5.5.5 Section sec_5.5.6 Figure fig_5 Section sec_5.5.7 |
19 | Section sec_5.5.8 Section sec_5.5.9 Section sec_5.6 Section sec_5.6.1 Section sec_5.6.2 Section sec_5.6.3 5.6 Alignment in X-ray photoelectron microscope/photoelectron imaging system |
20 | Section sec_5.7 Section sec_5.7.1 Section sec_5.7.2 Section sec_5.7.3 Section sec_5.7.4 Figure fig_6 5.7 Alignment by observing direct ion beam spot or crater image during and/or after ion sputtering |
21 | Section sec_5.8 Section sec_5.8.1 Section sec_5.8.2 Section sec_6 Section sec_6.1 Section sec_6.2 Section sec_6.3 5.8 Alignment by observing phosphor sample 6 When to align and check ion beam alignment |
22 | Annex sec_A Annex A (informative) Comparison of AES depth profiles with good/poor ion beam alignment |
23 | Figure fig_A.1 |
24 | Annex sec_B Annex sec_B.1 Annex sec_B.2 Table tab_b Figure fig_B.1 Annex B (informative) Alignment using cup with co-axial electrodes |
26 | Reference ref_1 Reference ref_2 Reference ref_3 Reference ref_4 Reference ref_5 Reference ref_6 Reference ref_7 Reference ref_8 Reference ref_9 Reference ref_10 Reference ref_11 Reference ref_12 Bibliography |